Ensure uptime with accurate, low-level sodium measurements and predictive diagnostics.

Flow Rate100 - 150 mL/min (6 - 9 L/h)
Analog outputs6 isolated, 0 - 20 mA or 4 - 20 mA; load impedance: 600 Ohm maximum. Connection: 0.644 - 1.29 mm² (24 - 16 AWG) wire; 0.644 - 0.812 mm² (24 - 20 AWG) twisted-pair shielded cable recommended
Relative humidity10 - 80%, non-condensing
Maintenance intervalEvery 90 days: refill of electrolytes, reagents and calibration solution
DisplayColour 5.7'' LCD
Electrode typeSodium ISE (ion-selective electrode) and reference electrode with KCl electrolyte
Power requirements (Voltage)100 - 240 V AC
InputSample line and sample bypass drain: 6 mm O.D. push-in fitting for plastic tubing. Chemical and case drain: 7/16 inch I.D. slip fitting for soft plastic tubing
Interference< 0.1 ppb
Calibration methodAutomatic, and Manual with standard addition technique: 1 or 2 points
Pollution degree2
MalzemePolyol case, PC door, PC hinges and latches, 304/316 SST hardware
Sample pressure0.2 - 6 bar
Max. suspended solids concentration in sample< 2 NTU, without oil and grease. Install an approx. 100 µm filter for boiler sample type
Sample temperature5 - 45°C
Ambient temperature5 - 50°C
ParameterSodium
Relay output6; type: unpowered SPDT relays, each rated 5 A resistive, 240 VAC maximum. Connection: 1.0 - 1.29 mm² (18 - 16 AWG) wire; 1.0 mm² (18 AWG) stranded wire recommended, 5 - 8 mm O.D. cable
Number of relays6
Storage conditions-20 - 60°C
Detection limit0.01 ppb
FuseInput power: T 1.6 A, 250 VAC; Relays: T 5.0 A, 250 V
Repeatability0.02 ppb or 1.5% of reading (whichever is greater) within a change of < ±10°C
Response time0.1 ppb - 10 ppb: ≤ 3 minutes, T95 ≤ 4 minutes; < 1 ppb - 100 ppb: T90 < 2 minutes, T95 < 3 minutes (approx. 150 seconds)
Total number of analog outputs6
Load600 Ohm