FX6x0 Sample Filtration Systems
The FX610 and FX620 are lightweight sample filtration systems for the NH6000sc and include a sample pump and compressor integrated directly into the analyser.
Simplify your online operations with the NH6000sc online Ammonium Analyser using gas-sensing electrode (GSE) technology. The climate-controlled design is ready for easy outdoor installation at the measurement point you need.
| Weight | Approx. 45 kg without chemicals |
| Flow Rate | 1.0 - 20.0 L/h from external sample filtration to the analyser |
| Boyutlar | 575 x 991 x 425 mm (22.63 x 39.01 x 16.73 in.) |
| Output | Relay, analog outputs, network interface via SC Controller |
| Accuracy | With Standard Solution: Measuring range 1: ≤1 mg/L: 3% of measured value + 0.02 mg/L; > 1 mg/L: 5% of measured value + 0.02 mg/L. Measuring range 2: 3% of measured value + 0.05 mg/L. Measuring range 3: 3% of measured value + 1.0 mg/L. Measuring range 4: 4.5% of measured value + 10 mg/L |
| Power supply | Power supplied by the SC controller. An additional power box is required with the SC4500. Analyser and heated drain line (powered by the analyser): 115 VAC or 230 VAC |
| Indoor/Outdoor | Indoor or outdoor use |
| Operating temperature range | -20 - 45°C; 95% relative humidity, non-condensing |
| Permitted chloride range | ≤1000 mg/L Cl - |
| Cable Length | Data cable length: < 50 m; Power cable length: < 5 m |
| Pollution degree | 2 |
| Controller compatibility | Digital SC Controller. Controllers sold separately. |
| Protection class | Class 1 |
| What's in the box | NH6000sc Ammonium Analyser, Mounting Hardware, User Manual |
| Malzeme | PUR 66, UV resistant |
| Enclosure Rating | Derece: IP55, NEMA UL50E 3R Malzeme: PUR 66 |
| Sample pressure | Maximum 2.5 bar (0.25 MPa) |
| Sample quality | Ultra-filtered or similar (FX6x0 < 0.45 µm) |
| Sample temperature | 4 - 40°C |
| Measuring range | Measuring range 1: 0.02 - 5.0 mg/L NH₄-N; Measuring range 2: 0.05 - 20 mg/L NH₄-N; Measuring range 3: 1 - 100 mg/L NH₄-N; Measuring range 4: 10 - 1000 mg/L NH₄-N |
| Measurement method | Gas sensing electrode (GSE) |
| Measurement frequency | 5, 10, 15, 20 or 30 min (user-adjustable) |
| Important notes | The liquid level in the basin must be below the bottom of the analyser |
| Parameter | Amonyum |
| pH Range | 5 - 9 pH |
| Storage conditions | -20 - 60°C; 95% relative humidity, non-condensing |
| Detection limit | Measuring range 1: 0.02 mg/L NH₄-N; Measuring range 2: 0.05 mg/L NH₄-N; Measuring range 3: 1 mg/L NH₄-N; Measuring range 4: 10 mg/L NH₄-N |
| Certifications | CE, UKCA, CMIM, FCC, ISED, certified to UL and CSA safety standards by TÜV |
| Repeatability | With Standard Solution: Measuring range 1: 3% of measured value + 0.02 mg/L; Measuring range 2: 2% of measured value + 0.05 mg/L; Measuring range 3: 2% of measured value + 1.0 mg/L; Measuring range 4: 2% of measured value + 10 mg/L |
| Response time | Measuring range 1: ≤0.2 mg/L: 3 measurements required (minimum 15 minutes in total); > 0.2 mg/L: 1 measurement required (5 minutes in total). Measuring range 2: < 5 minutes. Measuring range 3: < 5 minutes. Measuring range 4: < 5 minutes |
| Height | 2000 m maximum |
| Part No. | Description | Filter Configuration | Options | Number of channels |
|---|---|---|---|---|
| LXV461.99.00211 | NH6000sc Ammonium Analyser, 1 channel, for use with 1 external filtration system, flow detection, 230 VAC | Can be used e.g. with Filtrax. | Flow detection | 1 |
| LXV461.99.00311 | NH6000sc Ammonium Analyser, 1 channel, for use with 1 external filtration system, flow detection, automatic grab sampling, 230 VAC | Can be used e.g. with Filtrax. | Flow detection + automatic grab sampling | 1 |
| LXV461.99.01211 | NH6000sc Ammonium Analyser, 1 channel, integrated sample feed pump for FX610 filtration unit, flow detection, includes 230 VAC | Integrated sample feed pump for FX610 filtration unit | Flow detection | 1 |
| LXV461.99.01311 | NH6000sc Ammonium Analyser, 1 channel, sample feed pump for FX610, flow detection, automatic grab sampling, includes 230 VAC | Integrated sample feed pump for FX610 filtration unit | Flow detection + automatic grab sampling | 1 |
| LXV461.99.02211 | NH6000sc Ammonium Analyser, 1 channel, sample feed pump for use with FX620 filtration unit, flow detection, includes 230 VAC | Integrated sample feed pump for FX620 filtration unit | Flow detection | 1 |
| LXV461.99.02311 | NH6000sc Ammonium Analyser, 1 channel, sample feed pump for FX620, flow detection, automatic grab sampling, includes 230 VAC | Integrated sample feed pump for FX620 filtration unit | Flow detection + automatic grab sampling | 1 |
| LXV461.99.03211 | NH6000sc Ammonium Analyser, 2 channels, for use with 2 external filtration systems, flow detection, 230 VAC | Can be used e.g. with Filtrax. | Flow detection | 2 |
| LXV461.99.04211 | NH6000sc Ammonium Analyser, 2 channels, for use with 1 external sample filtration unit, with 1 sample pump for FX610, flow detection, 230 VAC | Integrated sample feed pump for FX610 filtration unit | Flow detection | 2 |
| LXV461.99.05211 | NH6000sc Ammonium Analyser, 2 channels, for use with 1 external sample filtration unit, with 1 sample pump for FX620, flow detection, 230 VAC | Integrated sample feed pump for FX620 filtration unit | Flow detection | 2 |
Request a quote for pricing and lead time; you can mention the part number you are interested in your message.
Documents open on the Hach website in a new tab.
Simplify your online operations with the NH6000sc online Ammonium Analyser using gas-sensing electrode (GSE) technology. The climate-controlled design is ready for easy outdoor installation at the measurement point you need. With minimal maintenance requirements and proven reliability, the Hach NH6000sc gives you confidence in measurement results and improved process efficiency.
Maximise uptime and accuracy with reliable technology
The Hach NH6000sc ammonium analyser uses fast, accurate and reliable GSE technology. The NH6000sc provides reliable measurements with automatic calibration, verification and cleaning. A grab verification sample improves consistency between online and laboratory data through an automatic sampling process. Predictive diagnostics and supportive maintenance workflows help you maximise uptime.
Save time with maintenance twice a year
The NH6000sc ammonium analyser reduces routine user maintenance to just twice a year. Designed with efficiency in mind, its modern design and optimised reagents make it simple and user-friendly. With the standard cleaning cycle, the reagent replacement interval has been extended to once every 6 months.
Robust, lightweight integrated filtration options
The integrated FX610 filtration system and the more robust FX620 offer a lightweight design with advanced air cleaning and innovative flow detection. These features provide reliable functionality and fewer operational touchpoints. You can also remotely check whether your sample is continuously being delivered to your analyser to increase the reliability of your results.
The NH6000sc, the new and improved version of the Amtax sc ammonium analyser, offers enhanced performance and reliability for your ammonium analysis needs.