Hach

Polymetron 2586sc Oxygen Scavenger Analyser

An integral part of the most complete water analysis system designed for the power industry. Hach provides a wide range of products designed to work together to build flexible solutions that meet your unique requirements.

Weight14.6 kg
Flow166 - 250 mL/min (10 - 15 L/h) recommended
Relative humidity10 - 90% relative humidity (non-condensing)
Connection4 x 6 mm stainless steel tubing
Maintenance intervalMonthly: Calibration and reagent refill
Pressure range0.5 - 6 bar (7.2 - 87 psi) or 12 L/h
Drain line connection6 x 8 mm (tubing must not exceed 4 ft and must drain straight down)
Electrical certificationsEMC: CE compliant for conducted and radiated emissions: - CISPR 11 (Class A limits) - EMC Immunity EN 61326-1 (Industrial limits). Safety: CAN/CSA C22.2 No. 61010-1; cETLus safety mark: - General Locations per ANSI/UL 61010-1 and CAN/CSA C22.2 No. 61010-1
Safety2
Communication (optional)Analog: five 0 - 20 mA or 4 - 20 mA analog outputs per analog output module; up to two analog Input modules (0 - 20 mA or 4 - 20 mA). Each input module replaces one digital sensor input. Digital: Profibus DPV1 module, Modbus TCP, Profinet IO module, Ethernet IP module
Operating temperature range5 - 45°C at 0 - 95% relative humidity (non-condensing)
Calibration methodSlope: using a laboratory reference value (e.g. LCW025)
Number of channels1
MalzemeCounter Electrode: Stainless steel; Working Electrode: Platinum
Manual languagesBulgarian, Chinese (PRC), Croatian, Czech, Danish, Dutch, English, Estonian, Finnish, French, German, Greek, Hungarian, Italian, Japanese, Korean, Lithuanian, Polish, Portuguese (Brazil), Portuguese (Portugal), Romanian, Russian, Slovak, Slovenian, Spanish, Swedish, Thai, Turkish
Enclosure RatingUL50E type 4X, IEC/EN 60529–IP 66, NEMA 250 type 4X. Metal enclosure with corrosion-resistant finish
Sample requirementsThe sample must not contain undissolved matter.
Sample temperature5 - 45°C
Measuring range0 - 500 ppb hidrazin; programlanabilir
Measuring range (2)0 - 100 ppb carbohydrazide (also known as ELIMIN-OX); programmable. ELIMIN-OX is a registered trademark of Nalco Chemical Co., Naperville, IL.
ReagentSample conditioning with diisopropylamine, ammonia or diethylamine
ReferansAg/AgCl electrode with 0.1 M KCl electrolyte
Relay: Operating modePrimary or secondary measurement or timer/scheduler
Relay functionsScheduler (Timer), Alarm, Feeder Control, Event Control, Pulse Width Modulation, Frequency Control and Warning
RelaysFour electromechanical SPDT (Form C) contacts, 1200 W, 5 A
Storage conditions-20 - 70°C, 0 - 95% relative humidity, non-condensing
Detection limitDrift negligible; 1 ppb
Repeatability±2% or 1 ppb (whichever is greater)
Response time< 60 s
Compliance certificationsCE. ETL certified to UL and CSA safety standards (all sensor types), FCC, ISED, KC, RCM, EAC, UKCA, SABS, C (Morocco)