Hach
SP510 Hardness Analyser
Maximise softener cycle time and minimise regeneration costs. Signals the hardness limit immediately to trigger regeneration. Makes your water softening systems more efficient and less costly.
| Weight | 11.3 kg |
| Flow Rate | 50 - 500 mL/min |
| Alarm | 2 consecutive cycles above the set point activate the alarm. 1 cycle below the set point cancels the alarm. Alarm type: LED indicators, Hard or Soft |
| Fitting | Inlet: ¼ inch OD tubing quick-disconnect fitting; Outlet: ½ inch ID flexible tubing |
| Maintenance interval | Standardise and refill reagents every 2 months using EDTA or a high-level sample. Replace pump tubing every 6 months |
| Dimensions (H x W x D) | 178 mm mm x 419 mm mm x 318 mm |
| Output | 1 SPDT relay |
| Detector | Silicon photo detector |
| Accuracy | ±25% of set point value |
| Drenaj | ½ in. ID flexible hose |
| Power requirements (Hz) | 50/60 Hz |
| Power requirements (Voltage) | 115/230 VAC, 70 VA, 50/60 Hz, 1.25 A Fuse |
| Warranty | 24 aylar |
| Operating temperature range | 5 - 40 °C |
| What's in the box | SP510 Hardness Monitor, two-month reagent supply, an installation kit, a maintenance kit, Hach user manual. |
| Malzeme | ABS plastic, large plastic windows to view alarm and reagent level |
| Enclosure Rating | IP62 |
| Sample pressure | 0.07 - 0.34 bar (0.10 bar is optimum). Sample conditioning: 0.10 - 5.17 bar |
| Sample Inlet | ¼ in. OD tube quick-disconnect fitting |
| Measuring range | Hardness levels: 0.3, 1, 2, 5, 10, 20, 50, 100 ppm (expressed as mg/L CaCO3) |
| Measurement method | Colorimetric |
| Measurement time | Average 2.0 min. |
| Optical path length | 8.89 mm |
| Reagent consumption | 500 mL of each indicator and buffer every two months |
| Certifications | NRTL certified to UL and CSA standards and CE approved |
| Repeatability | ±10% of set point value in the 0.3 - 2 mg/L ranges; ± 4% of set point value in the 5 - 100 mg/L ranges |
| Part No. | Description |
|---|---|
| 5410001-EU | SP510 Hardness Analyser (1 mg/L trigger point) |
| 5410002-EU | SP510 Hardness Analyser (2 mg/L trigger point) |
| 5410005-EU | SP510 Hardness Analyser (5 mg/L trigger point) |
| 5410010-EU | SP510 Hardness Analyser (10 mg/L trigger point) |
| 5410020-EU | SP510 Hardness Analyser (20 mg/L trigger point) |
| 5410050-EU | SP510 Hardness Analyser (50 mg/L trigger point) |
| 5410099-EU | SP510 Hardness Analyser (100 mg/L trigger point) |
Request a quote for pricing and lead time; you can mention the part number you are interested in your message.
- Tender Text Hardness Monitor_DOC353.94.35086
- SP510 Hardness Monitor
- Care Advantage Brochure : Complete service solution for all your instruments
- Care Advantage Flyer: Complete service solution for all your instruments
- Complete Water Analysis for Chemical and Refining industry
- Complete Water Analysis Solutions for Power and Steam Generation
- Complete Water Analysis Solutions for Power and Steam Generation
- CAD Drawing for SP510 Hardness Monitor_SP510 CAD Drawing
- Tender Text Hardness Monitor_DOC353.52.35086
- SP510 Hardness Monitor
Documents open on the Hach website in a new tab.
Maximise softener cycle time and minimise regeneration costs.
Signals the hardness limit immediately to trigger regeneration. Makes your water softening systems more efficient and less costly. Reliable, simple and accurate with automatic calibration. "Real-time" continuous monitoring.