| Weight | 20 kg without reagents and standards; 36.3 kg with reagents |
| Flow Rate | 55 - 300 mL/min |
| Alarm | Four relay outputs; type: unpowered SPDT relays, each rated 5 A resistive, 240 VAC maximum. Connection: 18 to 14 AWG wire, 18 AWG stranded wire recommended. |
| Fitting | Sample line and sample bypass drain: 6 mm (¼-in.); Air purge inlet: 6 mm (¼-in.); Chemical and enclosure drains: 9.5 mm (3/8-in.) |
| Output | 4 - 20 mA (4x) |
| Accuracy | 0 - 500 µg/L: ±1 µg/L or ±1% of reading, whichever is greater; 500 - 5000 µg/L: ±5% |
| Warranty | 24 aylar |
| External sample | Grab sample in and grab sample out feature |
| Light Source | Class 1M LED (light emitting diode) with 810 nm peak wavelength |
| Operating humidity | 5 - 95% non-condensing (indoor use only) |
| Operating temperature range | 5 - 45 °C |
| Pollution degree | 2/II |
| Controller compatibility | SC200, SC1000 |
| Protection class | I |
| What's in the box | Hach 5500 sc Silica Analyser, reagent set for up to 90 days of use, installation kit, user manuals |
| Manual languages | English, French, Spanish, Brazilian Portuguese, Chinese, Japanese, Korean, Thai, German, Italian, Portuguese, Czech, Danish, Dutch, Polish, Swedish, Finnish, Bulgarian, Hungarian, Romanian, Lithuanian, Russian, Turkish, Slovak, Slovenian, Croatian, Greek, Estonian |
| Mounting | Wall, panel or benchtop |
| Enclosure Rating | IP56 / NEMA 4X |
| Sample pressure | 0.14 - 6 bar (for Pressure Regulator Preset) |
| Sample temperature | 5 - 50 °C |
| Measuring range | 0 - 5000 µg/L as SiO 2 |
| Measuring principle | Colorimetric |
| Parameter | Silica |
| Reagent consumption | 2 L reagent for 90 days with 15-minute measurement time |
| Relays | Four; type: unpowered SPDT relays, each rated 5 A resistive, 240 V AC maximum |
| Storage conditions | -20 - 60 °C |
| Detection limit | 0.5 µg/L |
| Certifications | CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 ) KC (EN 61326-1: 2006) C-tick (EN 61326-1: 2006) cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012) |
| Repeatability | ± 0.5 µg/L or ± 1% of reading, whichever is greater |
| Response time | Typically 9.5 minutes at 25°C; varies with temperature |
| Uygulama | Pure water / Power |