| Weight | Max. 35 kg |
| Flow Rate | 100 - 300 mL/min |
| Analog outputs | Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8 |
| Dimensions (H x W x D) | 688 mm x 460 mm x 340 mm |
| Output | Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP |
| Instrument air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
| Demineralised water | Internal dilution option: Rinse water & Dilution water |
| Digital outputs | Profibus DP or Modbus RTU |
| Verification | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Accuracy | Better than 2% full scale range for standard test solutions |
| Drenaj | Atmospheric pressure, vented, min. Ø 32 mm |
| Power | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
| Warranty | 24 aylar |
| Interferences | Ions, e.g., antimony (III) (Sb³⁺), bismuth (III) (Bi³⁺), chloroplatinate (PtCl₆²ˉ), gold (III) (Au³⁺), iron(III) (Fe³⁺), lead (II) (Pb²⁺), mercury (II) (Hg²⁺), metavanadate (VO₃ˉ) and silver (I) (Ag⁺) can cause precipitation with nitrate. Copper (Cu²⁺) ions can decompose the diazonium salt cause a low result. NCl₃ can cause a red colour. Strong oxidizing agents interfere. Large quantities of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere. |
| Calibration | Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced |
| Protection class | IP44 |
| What's in the box | EZ1029 Nitrate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x empty 2.5L Reagent Container with Fittings (for Sulfamic Acid & Buffer Solution), 1 x empty 5L Reagent Container with Fittings (for Reductor Solution) and 1 x empty 2.5L Glass Reagent Container with Fittings (for Colour Solution) |
| Malzeme | Hinged part: Thermoform ABS, door: PMMA Wall section: Galvanised steel, powder coated |
| Number of sample streams | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
| Sample pressure | By external overflow vessel (open to atmospheric pressure) |
| Sample quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Sample temperature | 10 - 30 °C |
| Measuring range | Internal Dilution 0.2 - 10 mg/L NO₃-N |
| Measurement method | Colorimetric
NEDD colorimetric method after reduction with hydrazine |
| Measurement time | Default: 20 min. Continuous: 16 min. |
| Ambient temperature | 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
| Automatic cleaning | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Parameter | Nitrate |
| Reagent requirement | Keep Reducting reagent and Color Reagent refrigerated: 10 - 15 °C; all other between 10 - 30 °C |
| Relays | 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) |
| Detection limit | 0.2 - 10 mg/L NO₃-N: 0.2 mg/L |
| Certifications | CE, ETL certified to UL and CSA safety standards, UKCA |
| Earth connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |