| Weight | Max. 35 kg |
| Flow Rate | 100 - 300 mL/min |
| Analog outputs | Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard: 4, optional: 8 |
| Dimensions (H x W x D) | 740 mm x 500 mm x 340 mm |
| Output | Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP |
| Instrument air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
| Demineralised water | Internal dilution option: Rinse water & Dilution water |
| Digital outputs | Profibus DP or Modbus RTU |
| Verification | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Accuracy | Better than 2% full scale range for standard test solutions |
| Drenaj | Atmospheric pressure, vented, min. Ø 32 mm |
| Power | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
| Warranty | 24 aylar |
| Interferences | Other metals that form complexes with EDTA interfere. Silica (SiO₂) ˃ 500 mg/L, organic matter in water. Suspended or colloidal organic matter also can interfere with the endpoint. Large quantities of colour and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere. |
| Calibration | Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced |
| Protection class | IP44 |
| What's in the box | EZ1036sc Sulphate Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x Assembly Container 5L (for Buffer & BaCl₂ Solution) and 1 x Assembly Container 2.5L (for EDTA Solution) |
| Malzeme | Hinged part: Thermoform ABS, door: PMMA Wall section: Galvanised steel, powder coated |
| Number of sample streams | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
| Sample pressure | By external overflow vessel (open to atmospheric pressure) |
| Sample quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Sample temperature | 10 - 30 °C |
| Measuring range | No Dilution 5 - 40 mg/L SO₄²⁻ Internal Dilution 25 - 200 mg/L SO₄²⁻ 50 - 400 mg/L SO₄²⁻ 125 - 1000 mg/L SO₄²⁻ 250 - 2000 mg/L SO₄²⁻ 375 - 3000 mg/L SO₄²⁻ 500 - 4000 mg/L SO₄²⁻ |
| Measurement method | Barium chloride precipitation colorimetric method
Colorimetric |
| Measurement time | No Dilution Default: 10 min. Continuous: 10 min. Internal Dilution Default: 15 min. Continuous: 13 min. |
| Ambient temperature | 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
| Automatic cleaning | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Parameter | Sulphate |
| Reagent requirement | Keep between 10 - 30 °C |
| Relays | 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) |
| Detection limit | No Dilution 5 - 40 mg/L SO₄²⁻: 5 mg/L Internal Dilution 25 - 200 mg/L SO₄²⁻: 25 mg/L 50 - 400 mg/L SO₄²⁻: 50 mg/L 125 - 1000 mg/L SO₄²⁻: 125 mg/L 250 - 2000 mg/L SO₄²⁻: 250 mg/L 375 - 3000 mg/L SO₄²⁻: 375 mg/L 500 - 4000 mg/L SO₄²⁻: 500 mg/L |
| Certifications | CE, ETL certified to UL and CSA safety standards, UKCA |
| Earth connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |