| Weight | Max. 35 kg |
| Flow Rate | 100 - 300 mL/min |
| Dimensions (H x W x D) | 688 mm x 460 mm x 340 mm |
| Output | Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP |
| Instrument air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
| Demineralised water | Internal dilution option: Rinse water & Dilution water |
| Digital outputs | Profibus DP or Modbus RTU |
| Verification | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Accuracy | Better than 2% full scale range for standard test solutions |
| Drenaj | Atmospheric pressure, vented, min. Ø 32 mm |
| Warranty | 24 aylar |
| Interferences | Arsenic (V) [(As)5+], chromium (VI) [(Cr)6+], copper (II) [(Cu)2+] > 10 mg/L, iron (III) [(Fe)3+] > 10 mg/L, sulfide [(S)2-] > 2 mg/L and vanadium [(V)5+], silica [(Si)4+] > 60 mg/L. Large quantities of color and turbidity interfere. Fats, oil, proteins, surfactants and tar interfere. |
| Calibration | Automatic; 2-point; offset or slope; Frequency freely programmable: 6 hours, 12 hours, daily, weekly. Note: manufacturer recommends that a calibration is done when the reagents are replaced |
| Protection class | IP44 |
| What's in the box | EZ2732sc Total Phosphorus Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets and empty Containers for Reagents with Fittings |
| Malzeme | Hinged part: Thermoform ABS, door: PMMA Wall section: Galvanised steel, powder coated |
| Model | EZ2732sc |
| Enclosure Rating | IP44 |
| Number of sample streams | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
| Sample pressure | By external overflow vessel (open to atmospheric pressure) |
| Sample quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Sample temperature | 10 - 30 °C |
| Measurement method | Colorimetric
Molybdate blue method conform with APHA 4500-P |
| Measurement time | No Dilution Default: 45 min. Continous: 31 min. Internal Dilution Default: 33 min. Continous 33 min. |
| Ambient temperature | 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
| Automatic cleaning | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Parameter | Phosphorus, total |
| Reagent requirement | Keep Reducing agent between 10 - 15 °C, all other between 10 - 30 °C |
| Relays | 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) |
| Certifications | CE, ETL certified to UL and CSA safety standards, UKCA |
| Earth connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |