| Weight | 46 kg |
| Range selection | Automatic or Manual Range Selection |
| Multi-Stream | Up to 2 process streams and grab sample reading |
| Languages | English, Spanish, French, German. Other languages available on request. |
| EExp / Hazardous Area | Certification options are available for European Standards (ATEX Zone 1, Zone 2), North American Standards (Class I Division 2) and IECEx Zone 1 |
| Display | High-contrast 40 characters x 16 lines backlit LCD with LED backlight |
| Power requirements (Hz) | 50 Hz |
| Power requirements (Voltage) | 230 V AC |
| Warranty | 12 aylar |
| Power supply | TIC/TOC - 230V |
| Communication: digital | Modbus RTU, Modbus TCP/IP & Profibus (when the Profibus option is selected, digital output signals are sent via a Profibus converter using a proprietary communication protocol). Except for Zone 1 certification, Modbus RTU, Modbus TCP/IP and Modbus TCP/IP Redundant are available |
| Operating temperature range | 5 - 40 °C |
| Permitted chloride range | En fazla %30 w/v |
| Protection class | IP44, standard fan cooled, maximum ambient temperature 45 °C; IP54, air cooled, maximum ambient temperature 35 °C; IP54, vortex cooled, maximum ambient temperature 50 °C |
| User Interface | Microcontroller with membrane keypad |
| What's in the box | Contents: B3500c analyser, Tubing, Fuses, Ferrules, Drain, Acid & Base Immersion Tubes, CO₂ Filter & B3500c User Manual |
| Nem | 5 - 85% (non-condensing) |
| Sample inlet temperature | 0 - 60 °C |
| Oxidation Method | Innovative Two-Stage Advanced Oxidation Process (TSAO) using Hydroxyl Radicals |
| Measurement method | Infrared CO₂ Measurement after Oxidation (DIN EN 1484:1997-08, ISO 8245:1999-03, EPA 415.1) |
| Measurement time | From 5.5 minutes, depending on range and application. |
| Ambient temperature | 5 - 45 °C |
| Parameter | Measures Total Organic Carbon, Total Inorganic Carbon, Total Carbon, Chemical Oxygen Demand and Biological Oxygen Demand by correlation, and Volatile Organic Carbon by calculation |
| Particle size | En fazla 100 µm |
| Service interval | 6-monthly service intervals |
| Data Storage | Previous 9999 reaction data |