All-Weather Samplers
The AS950 AWRS is a rugged, weatherproof sampler with an integrated refrigerator for applications that require temperature control to preserve samples.
IO9004 analog 0/4-20 mA outputs and inputs. The IO9004 offers 3 analog 0/4-20 mA outputs and 2 analog 0/4-20 mA inputs. The outputs transmit sample events or logged data to SCADA systems.
| Weight | 2.5 kg |
| Base type | N/A |
| Boyutlar | With enclosure fittings: 24.7 cm (9.71 in.) length, 30 cm (11.82 in.) width, 13 cm (5.12 in.) depth |
| Power requirements (Voltage) | N/A |
| Inputs | Enclosure cable entries: - High-voltage ports: 4-cable grommet (accommodates cable diameters 4.3 - 11.4 mm) - Low-voltage ports: 4-cable grommet (accommodates cable diameters 4.3 - 11.4 mm). User-installed ½ inch (16 mm) trade-size electrical conduit can be used on all ports |
| Power input | Powered by the controller |
| Operating conditions | Temperature: -40 to 50°C; Humidity: 100% (non-condensing) |
| Cable length | 1.22 m |
| Mounting | Fittings and hardware included. |
| Enclosure material | Polycarbonate (UL94 V0), hardware 18-8 SS |
| Enclosure Rating | IP66 / NEMA 4X |
| Sample container | Uygulanamaz |
| Probes included? | N/A |
| Rain Gauge Input | Uygulanamaz |
| Certifications | CETLus |
| Height | 2000 Metre |
| Part No. | Description | User connections | Analog inputs | Digital outputs | Analog outputs |
|---|---|---|---|---|---|
| 9494500 | IO9001 Module (connects via auxiliary port), includes 1 relay (high voltage) | None | None | ||
| 9494600 | IO9004 Module (connects via auxiliary port), includes multiple 0/4-20 mA outputs and inputs | (2) 0/4-20 mA inputs supported for mapping logged measurements to external processes (e.g. third-party ultrasonic level). Current input mode: • Accuracy: 0.5% of FSR over -40 to 50°C, 0.3% of FSR over -10 to 50°C • Resolution: (4) Low-voltage contact closures mapped to alarm events. • ±30 VDC max., 150 mADC • 20 VAC-RMS, 150 mAAC-RMS Wire Size: 24 AWG to 16 AWG solid or stranded copper (3) 0/4-20 mA outputs mapped to logged analog measurements (e.g. level, velocity, flow, pH) - can be loop-powered or self-powered. • Accuracy: 0.8% of FSR at -40 - 50°C, 0.4% of FSR at -10 - 50°C • Resolution: • External Loop Power Supply Voltage: 18 V max. • Internal Loop Power Supply Voltage: 14 V min., 18 V max. • Loop Minimum Resistance: 0 Ω • Load Voltage: 3.6 V max. at 25 mA Wire size: 24 AWG - 16 AWG solid or stranded copper |
Request a quote for pricing and lead time; you can mention the part number you are interested in your message.
Documents open on the Hach website in a new tab.
• IO9004 analog 0/4-20 mA outputs and inputs
The IO9004 offers 3 analog 0/4-20 mA outputs and 2 analog 0/4-20 mA inputs. The outputs transmit sample events or logged data to SCADA systems. The inputs can be used to bring external measurements, such as from an existing water quality sensor or flow meters, into the sampler application. The AS950's standard memory can log any of these inputs.
• IO9001 or IO9004 high-voltage relays and digital outputs
The IO9004 includes 4 high-voltage relay outputs, while the IO9001 includes one high-voltage relay output. These relays are used to indicate alarms or events and can also be programmed to switch AC mains line voltage to control higher electrical functions. Examples include a warning light or audible signal and switching a diversion valve or gate, or passing a control signal to another machine. In addition, the 4 digital outputs of the IO9004 provide event-based control to SCADA systems and PLCs.