| Ağırlık | Max. 35 kg |
| Akış Hızı | 100 - 300 mL/min |
| Analog çıkışlar | Active 0 - 20 mA (or 4 - 20 mA) max. 500 Ohm load, standard 4, optional: 8 |
| Boyut (Y x G x D) | 688 mm x 460 mm x 340 mm |
| Cihaz havası | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air. Used to flush the instrument in corrosive environement. Min. 0.2 bar - Max. 0.5 bar |
| Dijital çıkışlar | Relays: 5 contacts, not user configurable: Malfuntion, maintenance, analysis ready, sample ready, sample ready (EZ9150) Ethernet Connections: Optional: Claros Ethernet connection and Modbus TCP/IP Ethernet connector; LAN version; 10/100 Mbps or Profinet or Ethernet IP RS485 communication: Profibus DP or Modbus RTU |
| Doğrulama | Automatic; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Doğruluk | Better than 2% full scale range for standard test solutions |
| Drenaj | Atmospheric pressure, vented, min. Ø 32 mm |
| Enerji | 100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA |
| Garanti | 24 aylar |
| Girişimler | Metal ions like aluminium [(Al)³+] > 72 mg/L, calcium [(Ca)²+] > 108 mg/L and iron [(Fe)²+/(Fe)³+] > 150 mg/L. Fats, oil, proteins, surfactants and tar. |
| Koruma sınıfı | IP44 |
| Malzeme | Hinged part: Thermoform ABS, Door: PMMA Wall section: Galvanised steel, powder coated |
| Numune akışı sayısı | Standard: 1 Optional accessory for 2, 4, or 8 sample streams |
| Numune basıncı | By external overflow vessel (open to atmospheric pressure) |
| Numune kalitesi | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Numune sıcaklığı | 10 - 30 °C |
| Ortam sıcaklığı | 10 - 30 °C ± 4 °C deviation at 5 - 95% relative humidity (non-condensing) |
| Otomatik temizleme | Yes; Frequency freely programmable: 6 hours, 12 hours, daily, weekly |
| Parametre | Fluoride |
| Reaktif gereksinimi | Keep between 10 - 30 °C |
| Sertifikalar | CE, ETL certified to UL and CSA safety standards, UKCA |
| Toprak bağlantısı | Earth connection Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |