FX6x0 Sample Filtration Systems
The FX610 and FX620 are lightweight sample filtration systems for the NH6000sc and include a sample pump and compressor integrated directly into the analyser.
Less Maintenance, Less Downtime. Only two litres of reagent are needed for up to 90 days of analyser operation — twice as long as the 5000 Series.
| Weight | 20 kg without reagents and standards; 36.3 kg with reagents |
| Flow Rate | 55 - 300 mL/min |
| Alarm | Four relay outputs; type: unpowered SPDT relays, each rated 5 A resistive, 240 VAC maximum. Connection: 18 to 14 AWG wire, 18 AWG stranded wire recommended. |
| Fitting | Sample line and sample bypass drain: 6 mm (¼-in.); Air purge inlet: 6 mm (¼-in.); Chemical and enclosure drains: 9.5 mm (3/8-in.) |
| Output | 4 - 20 mA |
| Power requirements (Hz) | 50/60 Hz |
| Power requirements (Voltage) | 100 - 240 V AC |
| Warranty | 24 aylar |
| External sample | Grab sample in and grab sample out feature |
| Light Source | Class 1M LED (light emitting diode) with 810 nm peak wavelength |
| Operating humidity | 5 - 95% non-condensing (indoor use only) |
| Operating temperature range | 5 - 45 °C |
| Pollution degree | 2/II |
| Controller compatibility | SC200, SC1000 |
| Protection class | I |
| What's in the box | Hach 5500 sc Phosphate Analyser, reagent set for 90 days of use, installation kit |
| Manual languages | English, French, Spanish, Brazilian Portuguese, Chinese, Japanese, Korean, Thai, German, Italian, Portuguese, Czech, Danish, Dutch, Polish, Swedish, Finnish, Bulgarian, Hungarian, Romanian, Lithuanian, Russian, Turkish, Slovak, Slovenian, Croatian, Greek, Estonian |
| Mounting | Wall, panel or benchtop |
| Enclosure Rating | IP56 / NEMA 4X |
| Sample pressure | 0.14 - 6 bar (2 - 87 psi) (for Pressure Regulator Preset) |
| Sample temperature | 5 - 50 °C |
| Measuring principle | Colorimetric |
| Reagent consumption | 2 L reagent for 90 days with 15-minute measurement time |
| Relays | Four; type: unpowered SPDT relays, each rated 5 A resistive, 240 V AC maximum |
| Storage conditions | -20 - 60 °C |
| Certifications | CE (EN 61326-1: 2006; EN 61010-1: 2010; EN 60529: 1991, +A1:2000 ) KC (EN 61326-1: 2006) C-tick (EN 61326-1: 2006) cETLus (UL 61010-1: 2012; NEMA 250: 2003; CSA C22.2 No 61010-1: 2012) |
| Response time | Typically 9.5 minutes at 25°C; varies with temperature |
| Uygulama | Pure water / Power |
| Part No. | Description | Measuring range | Number of channels | Parameter | Reagents included | Power requirements (Voltage) |
|---|---|---|---|---|---|---|
| 5500.KTO.PH.A1E | 5500 sc Phosphate High Range Analyser, 1 channel, 100-240 VAC, reagents included | 200 - 50000 µg/L PO 4 olarak | 1 | Phosphate (HR) | Evet | 100 - 240 V AC |
| 5500.KTO.PH.A2E | 5500 sc Phosphate High Range Analyser, 2 channels, 100-240 VAC, reagents included | 200 - 50000 µg/L PO 4 olarak | 2 | Phosphate (HR) | Evet | 100 - 240 V AC |
| 5500.KTO.PH.A4E | 5500 sc Phosphate High Range Analyser, 4 channels, 100-240 VAC, reagents included | 200 - 50000 µg/L PO 4 olarak | 4 | Phosphate (HR) | Evet | 100 - 240 V AC |
| 5500.KTO.PL.A1E | 5500 sc Phosphate Low Range Analyser, 1 channel, 100-240 VAC, reagents included | 4 - 3000 µg/L PO 4 [Detection range 4 - 5000 µg/L as PO 4] | 1 | Phosphate (LR) | Evet | 100 - 240 V AC |
| 5500.KTO.PL.A2E | 5500 sc Phosphate Low Range Analyser, 2 channels, 100-240 VAC, reagents included | 4 - 3000 µg/L PO 4 [Detection range 4 - 5000 µg/L as PO 4] | 2 | Phosphate (LR) | Evet | 100 - 240 V AC |
Request a quote for pricing and lead time; you can mention the part number you are interested in your message.
Documents open on the Hach website in a new tab.
Less Maintenance, Less Downtime.
Only two litres of reagent are needed for up to 90 days of analyser operation — twice as long as the 5000 Series. The industry's only pressurised reagent delivery system eliminates frequent pump maintenance. Hach's proprietary Prognosys technology, with predictive diagnostic tools such as warning LEDs and high-visibility notification screens, helps you prevent unexpected downtime. No more struggling with tubes and straws and dripping reagents on the instrument, the floor or your clothes: simply find the colour-coded cap matching the sealed reagent bottle and give it a gentle twist. Grab Sample In and Grab Sample Out features allow a collected sample poured into the analyser to be analysed quickly and make it easy to take samples out of the analyser for verification by laboratory testing.