Hach · EZ7300.XXXXXXXX

EZ Series ATP Analysers

The EZ7300 Series Analysers offer online monitoring of microbial activity in water via ATP (adenosine triphosphate). Typical applications are cooling water, steam & power generation, drinking water and process water.

Weight30 kg
Flow Rate100 - 300 mL/min
Alarm1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts
Analog outputsActive 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option)
Dimensions (H x W x D)690 mm x 465 mm x 330 mm
Instrument airDry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air
Demineralised waterFor rinsing
Digital outputsOptional: Modbus (TCP/IP, RS485)
VerificationAutomatic; frequency freely programmable
AccuracyBetter than 4% full scale range for standard test solutions
DrenajAtmospheric pressure, vented, min. Ø 32 mm
Power100 - 240 VAC, 50/60 Hz Max. power consumption: 120 VA
Warranty24 aylar
InterferencesHigh concentrations of Hg 2+ , Cu 2+ , Zn 2+ , Cd 2+ , Fe 2+ . Total salt concentrations higher than 1 g/L. pH lower than 5.5 and higher than 8.
Power supply100 - 240 VAC, 50/60 Hz
CalibrationAutomatic, 2-point; frequency freely programmable
Protection classAnalyser cabinet: IP44 / Panel PC: IP65
What's in the boxEZ7300 ATP Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 2 x empty 5L Reagent Container with Fittings (NaOH & HCl Solution) and 3 x Screwcap/Fittings/Reference1/Reference2
MalzemeHinged part: Thermoform ABS, door: PMMA Wall section: Galvanised steel, powder coated
ModelEZ7300
Number of sample streams1, 2, 4, or 8
Sample pressureMaximum 3 bar for direct injection or by external overflow vessel via the grab sample pump
Sample temperature5 - 30 °C
Measuring range0.5 - 200 pg/mL
0.5 - 200 pg/mL
Measurement methodDetermination of adenosine triphosphate (ATP) by means of chemiluminescent reaction using luciferin and luciferase, conform with standard method ASTM D4012-81
Measurement time7 - 10 minutes incl. sample lysis
Ambient temperature7 - 23 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing)
ParameterATP
Reagent requirementKeep between 4 - 8 °C during operation, shielded from light. For prolonged storage, store at -20 °C in a dark environment.
Detection limit≤ 0.05 pg/mL (0.1 pM) ATP
CertificationsCE compliant / UL certified
Earth connectionDry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm²

Product code: EZ7300.XXXXXXXX

Related products

EZ Series Ammonium Analysers
Hach

EZ Series Ammonium Analysers

The Hach EZ Series Ammonium Analysers provide reliable, continuous online monitoring of ammonium and ammonia (NH₄/NH₃) across a wide range of water and wastewater applications, including municipal wastewater treatment, drinking water, …

Ammonium Analysers Details →
EZ Series Fluoride Analysers
Hach

EZ Series Fluoride Analysers

For drinking water, fluoridation systems, and industrial wastewater treatment, Industrial Fluoride Removal Process, the Hach EZ Series Fluoride Analysers provide reliable, continuous fluoride measurement using proven Ion Selective…

Fluoride Analysers Details →
EZ Series Phosphate Analysers
Hach

EZ Series Phosphate Analysers

The EZ sc Series Online Analysers offer multiple options to monitor Phosphate in water. Typical applications are wastewater, drinking water, steam & power generation, surface water.

Phosphate Analysers Details →
EZ Series Silica Analysers
Hach

EZ Series Silica Analysers

The EZ sc Series Online Analysers offer multiple options to monitor Silica in water. Typical applications are steam & power generation, surface water, process water. The EZ1000sc Series Analysers use colorimetric analysis to measure Silica.

EZ1035.9800140T Details →