| Weight | 25 kg |
| Flow Rate | 100 - 300 mL/min |
| Alarm | 1 x malfunctioning, 4 x user-configurable, max. 24 VDC/0.5 A, potential free contacts |
| Analog outputs | Active 4 - 20 mA max. 500 Ohm load, standard 1, max. 8 (option) |
| Dimensions (H x W x D) | 690 mm x 465 mm x 330 mm |
| Instrument air | Dry and oil free according to ISA-S7.0.01-1996 quality standard for instrument air |
| Demineralised water | For rinsing |
| Digital outputs | Optional: Modbus (TCP/IP, RS485) |
| Verification | Automatic; frequency freely programmable |
| Accuracy | Better than 5% full scale range for standard test solutions |
| Drenaj | Atmospheric pressure, vented, min. Ø 32 mm |
| Power | 230 VAC, 50/60 Hz 120 VAC, 50/60 Hz Max. power consumption: 440 VA |
| Warranty | 24 aylar |
| Interferences | Chloride > 1 g/L, inorganic reducing agents such as nitrites, sulphides and iron(II) will increase the result, aromatic hydrocarbons and pyridine are not oxidized to any appreciable extent. Some very volatile organic substances may escape the oxidation by evaporation. Straight chain aliphatic compounds are effectively oxidised by the silver sulphate/sulphuric acid solution. Fats, oil, proteins, surfactants and tar. |
| Calibration | Automatic, 2-point; frequency freely programmable |
| Protection class | Analyser cabinet: IP55 / Panel PC: IP65 |
| What's in the box | EZ7050 COD Analyser, Instruction Manual, 1 x Double Bit Door Key, 1 x Mounting Brackets, 1 x Combined Pt Electrode, Routine S7 Connection, 1 x empty 2.5L Reagent Container with Fittings (AgNO₃ Solution), 2 x empty 10L Reagent Containers with Fittings (Acid & Oxalate Solution) and 1 x empty 10L Reagent Dark Container with Fittings (KMNO₄ Solution) |
| Malzeme | Hinged part: Thermoform ABS, door: PMMA Wall section: Galvanised steel, powder coated |
| Number of sample streams | 1, 2, 4, or 8 |
| Sample pressure | By external overflow vessel |
| Sample quality | Maximum particle size 100 µm, < 0.1 g/L; Turbidity < 50 NTU |
| Sample temperature | 10 - 30 °C |
| Measuring range | 1 - 20 mg/L O 2
1 - 20 mg/L O₂ |
| Measurement method | Redox titration after oxidation by acid-potassium dichromate solution, conform with ISO 6060 method Or redox titration after oxidation by potassium permanganate solution, conform with ISO 8467 and JIS K0806 methods |
| Measurement time | 40 minutes, including oxidation time of 30 minutes. Remark: standard method for Cr destruction requires 120 minutes. |
| Ambient temperature | 10 - 30 °C ±4 °C deviation at 5 - 95% relative humidity (non-condensing) |
| Parameter | COD |
| Reagent requirement | Keep between 10 - 30 °C |
| Detection limit | ≤ 1 mg/L |
| Certifications | CE compliant / UL certified |
| Cooling water | Flow rate approx. 5 L/h; temperature max. 30 °C; pressure max. 0.5 bar |
| Earth connection | Dry and clean earth pole with low impedance (< 1 Ohm) using an earth cable of > 2.5 mm² |