Hach

TU5300sc/TU5400sc Online Laser Turbidimeters

With a unique optical design that sees much more of your sample than any other turbidimeter, the EPA-approved TU5 Series platform delivers the best accuracy and sensitivity at low levels while minimising measurement-to-measurement variability…

Weight2.7 kg (5.0 kg with all accessories)
Flow Rate100 - 1000 mL/min; optimum flow rate: 200 - 500 mL/min
FittingSample quick-connect fitting: ¼ in. for ¼ in. tubing
UnitNTU, FNU, TE/F, FTU, EBC
Dimensions (H x W x D)249 mm x 268 mm x 190 mm
Resolution0,0001 NTU / FNU / TE/F / FTU / EBC
VerificationLiquid: Stablcal, Formazin (0.1 - 40 NTU); Dry: Glass Rod < 0.1 NTU
Warranty24 aylar
Light SourceClass 2 laser product, integrated 650 nm (EPA, 0.43 mW) or Class 1 laser product, integrated 850 nm (ISO), max. 0.55 mW (compliant with IEC/EN 60825-1 and 21 CFR 1040.10 in accordance with Laser Notice No. 50)
Operating humidityRelative humidity: 5 - 95% at various temperatures, non-condensing
Operating temperature range0 - 50 °C
Cable Length1.6 m (5.25 ft), extendable up to 50 m (164 ft) for instrument without accessories
Stray Light<10 mNTU
Calibration methodFor Formazin and Stablcal: at 20 NTU for the 0 - 40 NTU range; at 20 NTU and 600 NTU for the full range. Maximum 6-point custom user calibration
ControllerSensor Only
Controller compatibilitySC4500, SC1000, SC200
MalzemeASA Luran S 777K / RAL7000, TPE RESIN Elastocon STK40, Thermoplastic Elastomer TPS-SEBS
Mounting configurationsDuvara Montaj
Enclosure RatingElectronics compartment IP55; all other functional units IP65 when the process cap/ACM is fitted to the TU5300sc / TU5400sc
Sample pressureMaximum 6 bar, against air in the sample temperature range 2 - 40°C
Sample temperature2 - 60 °C
RegionGlobal (Sensor Only)
Storage conditions-40 - 60°C
CertificationsCE compliant. US FDA accession number: 1420493-000 EPA version, 1420492-000 ISO version. Compliant with IEC/EN 60825-1 and 21 CFR 1040.10 (in accordance with Laser Notice No. 50). Australian ACMA Mark

Related products